Test Materials: ISO 7500-1 specifies the use of certified calibration standards, such as reference weights or load cells, traceable to national or international standards.
Test Process: The machine is calibrated by applying known loads and comparing the machine’s readings to reference standards, ensuring accuracy within specified tolerances.
Test Result Interpretation: Results are compared with manufacturer specifications and tolerance limits. Any discrepancies require recalibration to ensure the machine meets ISO 7500-1 accuracy